Safety Auxiliary Feedback Element for the Artificial Pancreas in Type 1 Diabetes.

Compartir
Autores UPV

Año

Revista

IEEE Transactions on Biomedical Engineering

Abstract

The artificial pancreas aims at the automatic deliv-ery of insulin for glycemic control in patients with type 1 diabetes, i.e., closed-loop glucose control. One of the challenges of the arti-ficial pancreas is to avoid controller overreaction leading to hypo-glycemia, especially in the late postprandial period. In this study, an original proposal based on sliding mode reference conditioning ideas is presented as a way to reduce hypoglycemia events induced by a closed-loop glucose controller. The method is inspired in the intuitive advantages of two-step constrained control algorithms. It acts on the glucose reference sent to the main controller shaping it so as to avoid violating given constraints on the insulin-on-board. Some distinctive features of the proposed strategy are that 1) it provides a safety layer which can be adjusted according to medical criteria; 2) it can be added to closed-loop controllers of any nature; 3) it is robust against sensor failures and overestimated prandial insulin doses; and 4) it can handle nonlinear models. The method is evaluated in silico with the ten adult patients available in the FDA-accepted UVA simulator.